Test Catalog

Test Id : DHPVB

Human Papillomavirus (HPV) Types 18, 31, 33, and 35, Droplet Digital PCR, Blood

Test Catalog

Useful For
Suggests clinical disorders or settings where the test may be helpful

Aid in the monitoring of cancers caused by human papillomavirus (HPV) high-risk genotypes 18, 31, 33, and 35

 

Disease monitoring after treatment of an HPV-related cancer and/or post-treatment surveillance as a prognostic indicator of cancer recurrence

 

This test should not be used to screen asymptomatic patients.

Method Name
A short description of the method used to perform the test

Droplet Digital Polymerase Chain Reaction (ddPCR)

NY State Available
Indicates the status of NY State approval and if the test is orderable for NY State clients.

No

Reporting Name
Lists a shorter or abbreviated version of the Published Name for a test

HPV-18/31/33/35, cfDNA, ddPCR, B

Aliases
Lists additional common names for a test, as an aid in searching

HPV

HPV-18

HPV-31

HPV-33

HPV-35

Human Papillomavirus

Cell free DNA

cell-free DNA

cfDNA

Circulating tumor cells

circulating tumor DNA

Specimen Type
Describes the specimen type validated for testing

Whole blood

Ordering Guidance

This test should be ordered to monitor circulating cell free DNA for the high-risk human papillomavirus (HPV)-18, -31, -33 or -35 genotypes in patients with a previous diagnosis. If the intent of the order is a pre-diagnostic screen for which the genotype is not previously known, order DHPVR / Human Papillomavirus (HPV) Type 16 with Type 18, 31, 33, and 35 Reflex, Droplet Digital PCR, Blood which can detect and quantify circulating cell free DNA from 5 high-risk human papillomavirus genotypes: HPV-16, -18, -31, -33 and -35.

 

For routine cervical cancer screening, order VHPV / Human Papillomavirus (HPV) Vaginal Detection with Genotyping for High-Risk Types by PCR.

Specimen Required
Defines the optimal specimen required to perform the test and the preferred volume to complete testing

Supplies: Streck Tan Top Tube Kit (T715)

Container/Tube: Streck Cell-Free DNA blood collection kit

Specimen Volume: Two 10-mL Streck Cell-Free DNA blood collection tubes

Additional Information: Only blood collected in Streck Cell-Free DNA tubes will be accepted for analysis.

Specimen Minimum Volume
Defines the amount of sample necessary to provide a clinically relevant result as determined by the testing laboratory. The minimum volume is sufficient for one attempt at testing.

One 10 mL Streck Cell-Free DNA blood collection tube

Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected

Specimen collected in tube other than Streck Cell-Free DNA collection tube Reject
  All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.

Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the performing laboratory, alternate acceptable temperatures are also included

Specimen Type Temperature Time Special Container
Whole blood Ambient (preferred) 7 days
Refrigerated 7 days

Useful For
Suggests clinical disorders or settings where the test may be helpful

Aid in the monitoring of cancers caused by human papillomavirus (HPV) high-risk genotypes 18, 31, 33, and 35

 

Disease monitoring after treatment of an HPV-related cancer and/or post-treatment surveillance as a prognostic indicator of cancer recurrence

 

This test should not be used to screen asymptomatic patients.

Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

Human papillomavirus (HPV) is a double-stranded DNA virus from the Papillomaviridae family. There are over 200 known types of HPV, which can be classified into low-risk types, associated with benign conditions such as genital warts, and high-risk types, which are linked to the development of cancers, including cervical, anal, oropharyngeal, vulvar, penile, and vaginal cancers.(1-3)

 

Human papillomavirus infects the basal epithelial cells of the skin or mucous membranes. The virus integrates into host epithelial cells, where it can exist as either an episomal (extrachromosomal) or integrated form in the host DNA.(4,5) Low-risk HPV types (eg. HPV-6 and -11) usually remain in episomal form, while high-risk types (eg. HPV-16 and -18) are more likely to integrate into the host genome, causing dysregulation of the cell cycle. The virus produces proteins (E6 and E7) that interfere with tumor suppressor genes, p53 and pRb, and may lead to uncontrolled cell proliferation and a higher risk of progression to malignancy.(5)

 

Most HPV infections are transient and cleared by the host immune system within 1 to 2 years. However, persistent infection with high-risk HPV types increases the likelihood of progression to premalignant lesions and ultimately invasive cancer.(5)

 

Human papillomavirus cell free DNA has been shown to be an important biomarker circulating in patients with HPV-related cancers, including those of the head and neck. Fragmented HPV is released into circulation by apoptotic tumor cells and can be detected by molecular techniques. This assay utilizes droplet digital polymerase chain reaction to detect, subtype, and quantify high-risk HPV cfDNA in plasma samples.

Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

Not detected

Interpretation
Provides information to assist in interpretation of the test results

A qualitative result will be provided for each of the four high-risk human papillomavirus (HPV) genotypes (18, 31, 33 and 35) detected by this assay.

 

A result of "Not Detected" indicates the absence of HPV-18, -31, -33 and/or -35 cfDNA in the plasma (see Cautions below).

 

A result of "Detected" indicates the presence of HPV-18, -31, -33 and/or -35 cfDNA in the plasma. If HPV-18, -31, -33 and/or -35 cfDNA is detected, a quantitative value will be reported in fragments/mL.

 

The quantification range of this assay is 100 to 309,600 fragments/mL.

 

A result of "<100 fragments/mL" indicates that HPV-18, -31, -33 and/or -35 cfDNA is detected in the plasma, but the assay cannot accurately quantify the HPV-18, -31, -33 and/or -35 cfDNA present below this level.

 

A result of ">309,600 fragments/mL" indicates that HPV-18, -31, -33 and/or -35 cfDNA is detected in the plasma, but the assay cannot accurately quantify the HPV-18, -31, -33 and/or -35 cfDNA present above this level.

 

Samples with HPV signal below the limit of detection but above the limit of blank will be reported as "not detected" with an accompanying comment: "This specimen yielded signal above the limit of blank but below the assay's limit of detection, which may not be reproducible and should be interpreted in the context of the patient’s clinical presentation. If clinically indicated, recollect and submit a new specimen for testing in 2 to4 weeks."

Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

This assay only detects genetic fragments of the following HPV genotypes: 18, 31, 33 and 35.

 

This assay will not detect other human papillomavirus (HPV) genotypes.

 

This assay will not specify an anatomic location or source of detectable levels of HPV.

 

A result of “Not Detected” does not necessarily eliminate the possibility of HPV infection, or a tumor caused by HPV. Serial measurement of HPV in platelet poor plasma may be recommended based on clinical presentation.

 

This assay should only be used for patients with a clinical history and/or symptoms consistent with a confirmed or suspected HPV-related tumor.

 

Results must be interpreted in the context of the patient’s clinical picture.

 

Variants within the HPV E6 and E7 gene regions targeted by this assay may affect primer and/or probe binding, resulting in the under quantitation of viral fragments or failure to detect the presence of viral fragments.

Supportive Data

Analytical Sensitivity/Detection Capability:

The 95% limit of detection, by genotype, for this assay are:

19.45 fragments/mL of human papillomavirus (HPV)-18 cfDNA

17.83 fragments/mL of HPV-31 cfDNA

9.20 fragments/mL of HPV-33 cfDNA

7.71 fragments/mL of HPV-35 cfDNA

 

These were determined following a serial dilution of plasma spiked with a known concentration of each HPV genotype, with the limit of detection determined through probit analysis.

 

Accuracy:

Forty-eight contrived samples yielded 100% overall, positive and negative agreement.

 

Precision:

Intra- and inter-assay reproducibility experiments yielded 100% qualitative precision. Replicates at three levels within the analytical measuring range produced quantitative values with standard deviations of 0.00 – 0.07 log[10] fragments/mL.

 

Analytical Specificity:

A panel of 38 bacterial, fungal, parasitic and viral nucleic acid extracts from related organisms or organisms commonly found in plasma were tested by this assay with no cross-reactivity detected.

 

Inclusivity (in silico analysis) was performed using the NCBI GenBank database. Homology of assay primers and probes was determined using 25 representative strains from all four high-risk HPV genotypes detected by this assay (HPV-18, -31, -33, 35). Homology of oligos ranged from 88.9% to 100% across all strains analyzed.

 

Reference Range:

Platelet poor plasma from 20 healthy donors was tested by this assay. No samples yielded detectable HPV-18, -31, -33, nor -35 cfDNA above the limits of detection.

Clinical Reference
Recommendations for in-depth reading of a clinical nature

1. HPV and Cancer. National Cancer Institute. Updated May 9, 2025. Accessed October 14,2025. Available at www.cancer.gov/about-cancer/causes-prevention/risk/infectious-agents/hpv-and-cancer

2. Jamal Z, Anjum F. Oropharyngeal Squamous Cell Carcinoma. In: StatPearls [Internet]. Updated April 27, 2023. Accessed October 14, 2025. Available at www.ncbi.nlm.nih.gov/books/NBK563268/

3. Doorbar J, Quint W, Banks L, et al. The biology and life-cycle of human papillomaviruses. Vaccine. 2012;30 Suppl 5:F55–F70. doi:10.1016/j.vaccine.2012.06.083

4. Gribb JP, Wheelock JH, Park ES. Human papilloma virus (HPV) and the current state of oropharyngeal cancer prevention and treatment. Dela J Public Health. 2023;9(1):26-28. doi: 10.32481/djph.2023.04.008

5. Munger K, Phelps WC, Bubb V, Howley PM, Schlegel R. The E6 and E7 genes of the human papillomavirus type 16 together are necessary and sufficient for transformation of primary human keratinocytes. J Virol. 1989;63(10):4417-4421

6. Fernholz EC, Routman DM, Van Abel KM, et al. Detection, quantitation, and genotyping of human papillomavirus circulating tumor DNA by droplet digital PCR. J Clin Microbiol. 2025;63(9):e0058525. doi:10.1128/jcm.00585-25

Method Description
Describes how the test is performed and provides a method-specific reference

This laboratory-developed test utilizes whole blood samples collected in Streck Cell-Free DNA blood collection tubes. Cell-free DNA (cfDNA) is isolated from double-centrifuged plasma using the EZ2 Connect extraction system (Qiagen). Nucleic acid eluate is combined with polymerase chain reaction (PCR) reaction mix and partitioned into approximately 20,000 droplets by oil emulsion on the Automated Droplet Generator (Bio-Rad). Following PCR on the VeritiPro Thermal Cycler (Applied Biosystems), droplets are analyzed for presence or absence of amplified viral target sequences using the QX600 Droplet Reader (Bio-Rad).

 

The PCR reaction mix contains two primer and probe sets directed to the E6 and E7 gene regions of the human papillomavirus (HPV) type 18 genome, as well as single primer and probe sets for genotypes 31, 33 and 35. Amplicons range in size from 65 to 106 base pairs. The reaction mix also contains a primer-probe set directed to the human reference gene (ribonuclease P protein subunit 30 [RPP300]), which serves as an internal process control. Each hydrolysis probe is coupled with unique fluorescent dye that can be differentiated on the QX600. Each droplet is individually tallied and evaluated for fluorescence. Results are analyzed using dedicated software and Poisson statistics to provide absolute quantification of target and reference concentrations. Calculated results are reported as detected, not detected or indeterminate for HPV-18, -31, -33, and/or -35 cfDNA. A concentration is also reported for samples with detectable levels (fragments of HPV cfDNA per milliliter of plasma).(Unpublished Mayo method)

PDF Report
Indicates whether the report includes an additional document with charts, images or other enriched information

No

Day(s) Performed
Outlines the days the test is performed. This field reflects the day that the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time before the test is performed. Some tests are listed as continuously performed, which means that assays are performed multiple times during the day.

Varies

Report Available
The interval of time (receipt of sample at Mayo Clinic Laboratories to results available) taking into account standard setup days and weekends. The first day is the time that it typically takes for a result to be available. The last day is the time it might take, accounting for any necessary repeated testing.

5 to 10 days

Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

7 days

Performing Laboratory Location
Indicates the location of the laboratory that performs the test

Mayo Clinic Laboratories - Rochester Main Campus
CLIA Number: 24D0404292

Fees :
Several factors determine the fee charged to perform a test. Contact your U.S. or International Regional Manager for information about establishing a fee schedule or to learn more about resources to optimize test selection.

  • Authorized users can sign in to Test Prices for detailed fee information.
  • Clients without access to Test Prices can contact Customer Service 24 hours a day, seven days a week.
  • Prospective clients should contact their account representative. For assistance, contact Customer Service.

Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR) product.

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Clinic Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

CPT codes are provided by the performing laboratory.

87799

LOINC® Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the order and results codes of this test. LOINC values are provided by the performing laboratory.

Test Id Test Order Name Order LOINC Value
DHPVB HPV-18/31/33/35, cfDNA, ddPCR, B In Process
Result Id Test Result Name Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
622876 HPV-18 ddPCR Result 61373-7
622877 HPV-18 ddPCR Quant In Process
622878 HPV-31 ddPCR Result 61375-2
622879 HPV-31 ddPCR Quant In Process
622880 HPV-33 ddPCR Result 61376-0
622881 HPV-33 ddPCR Quant In Process
622882 HPV-35 ddPCR Result 61377-8
622883 HPV-35 ddPCR Quant In Process

Test Setup Resources

Setup Files
Test setup information contains test file definition details to support order and result interfacing between Mayo Clinic Laboratories and your Laboratory Information System.

Excel | Pdf

Sample Reports
Normal and Abnormal sample reports are provided as references for report appearance.

Normal Reports | Abnormal Reports

SI Sample Reports
International System (SI) of Unit reports are provided for a limited number of tests. These reports are intended for international account use and are only available through MayoLINK accounts that have been defined to receive them.

SI Normal Reports | SI Abnormal Reports