Web: | mayocliniclabs.com |
---|---|
Email: | mcl@mayo.edu |
Telephone: | 800-533-1710 |
International: | +1 855-379-3115 |
Values are valid only on day of printing. |
Affinity-purified Sm antigens are bound to polystyrene microspheres, which are impregnated with fluorescent dyes to create a unique fluorescent signature. Sm antibodies, if present in diluted serum, bind to the Sm antigens on the microspheres. The microspheres are washed to remove extraneous serum proteins. Phycoerythrin (PE)-conjugated antihuman IgG antibody is then added to detect IgG anti-Sm bound to the microspheres. The microspheres are washed to remove unbound conjugate, and bound conjugate is detected by laser photometry. A primary laser reveals the fluorescent signature of each microsphere to distinguish it from microspheres that are labeled with other antigens, and a secondary laser reveals the level of PE fluorescence associated with each microsphere. Results are calculated by comparing the median fluorescence response for Sm microspheres to a 4-point calibration curve.(Package insert: Bioplex 2200 ANA Screen. Bio-Rad Laboratories, Hercules, CA 11/2011)
Monday through Saturday; 4 p.m.