This assay is performed on the Instrumentation Laboratory ACL TOP using the Diagnostica Stago LIATEST AT III kit. Antithrombin antigen is determined using automated latex immunoassay (LIA) methodology. Patient plasma, containing antithrombin antigen, is combined with a latex reagent containing rabbit antihuman antibodies. An antigen-antibody reaction takes place, causing the latex particles to agglutinate and form aggregates. The aggregates form diameters greater than the wavelength of the light (405 nm) passing through causing absorption of the light. This change in absorption is measured over time and reported as delta optical density. The increase in absorption is proportional to the concentration of antithrombin antigen present in the sample.(Package insert: LIATEST AT III. Diagnostica Stago S.A.S., Asnieres sur Seine, France, Rev. 02/2015)