Test ID: RFAMA    
Renal Function Panel, Serum

Method Description Describes how the test is performed and provides a method-specific reference

KS, NAS, CL:

Ion-selective electrode (ISE) (indirect potentiometry). The ISE module performs indirect measurement of electromotive force (EMF). The ISE module measures the EMF difference between an ISE and a reference electrode. The EMF of the ISE is dependent on the ion concentration of the sample. The EMF of the reference electrode is constant. An electronic calculation circuit converts EMF of the sample to the ion concentration of the sample.(Package insert: Roche Diagnostics ISE reagent; Indianapolis, IN, 2006)

 

Bicarbonate (HCO3):

This is a photometric rate reaction. Bicarbonate (HCO3-) reacts with phosphoenolpyruvate (PEP) in the presence of phosphoenolpyruvate carboxylase (PEPC) to produce oxaloacetate and phosphate. The oxaloacetate produced is coupled with NADH in the presence of malate dehydrogenase (MDH) to produce malate and NAD. The consumption of NADH causes a decrease in absorbance and is monitored in the UV range of 320 nm to 400 nm. The rate of change is directly proportional to the concentration of bicarbonate.(Package insert: Roche Bicarbonate reagent, Indianapolis, IN, July 2000)

 

AGAP:

This is a calculated result. The following equation is used to calculate the anion gap (A gap):

A gap = Na - (Cl + HCO3)

 

BUN:

This is a kinetic ultraviolet assay where urease cleaves urea to form ammonia and CO2. The ammonia formed then reacts with a-ketoglutarate and NADH in the presence of urease/glutamate dehydrogenase (GLDH) to yield glutamate and NAD. The decrease in absorbance, due to the consumption of NADH, is measured kinetically and is proportional to the amount of urea in the sample.(Package insert: Roche Urea/BUN reagent; Indianapolis, IN, Sept 2000)

 

CRTS1:

This enzymatic method is based on the conversion of creatinine with the aid of creatininase, creatinase, and sarcosine oxidase to glycine, formaldehyde, and hydrogen peroxide. Catalyzed by peroxidase, the liberated hydrogen peroxide reacts with 4-aminophenazone and hydroxy(tosyloxy)iodobenzene (HTIB) to form a quinone imine chromogen. The color intensity of the quinone imine chromogen formed is directly proportional to the creatinine concentration in the reaction mixture.(Package insert: Roche Diagnostics, Indianapolis IN, 12/2016)

 

CA:

Calcium ions react with 5-nitro-5'-methyl-BAPTA (NM-BAPTA) under alkaline conditions to form a complex. This complex reacts in the second step with EDTA. The change in absorbance is directly proportional to the calcium concentration and is measured photometrically.(Package insert: Roche Calcium Gen.2 reagent, Roche Diagnostic Corp, Indianapolis, IN, 2012 July)

 

GLURA:

Glucose in the serum, in the presence of hexokinase, is converted to glucose-6-phosphate (G-6-P). Glucose-6-phosphate dehydrogenase (G-6-PDH), in the presence of NADP, oxides G-6-P to gluconate-6-phosphate and NADPH. The rate of NADPH formation is directly proportional to glucose concentration in the serum and is measured photometrically.(Package insert: Roche Glucose Reagent. Indianapolis, IN, January 2000)

 

ALB:

The dye, bromcresol green (BCG), is added to serum in an acid buffer. The color intensity of the blue-green albumin-BCG complex is directly proportional to the albumin concentration and is determined photometrically.(Package insert: Roche Albumin reagent; Roche Diagnostic Corp., Indianapolis, IN, July 1999)

 

PHOS:

The method is based on the reaction of phosphate with ammonium molybdate to form ammonium phosphomolybdate (without reduction). The addition of an accelerator gives rise to a more rapid rate of reaction. Sample blanking yields more precise results.(Package insert: Roche Phosphorus reagent, Roche Diagnostics Corp, Indianapolis, IN, 1999)

PDF Report Indicates whether the report includes an additional document with charts, images or other enriched information

No

Day(s) and Time(s) Test Performed Outlines the days and times the test is performed. This field reflects the day and time the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time required before the test is performed. Some tests are listed as continuously performed, which means assays are performed several times during the day.

Monday through Sunday; Continuously

Analytic Time Defines the amount of time it takes the laboratory to setup and perform the test. This is defined in number of days. The shortest interval of time expressed is "same day/1 day," which means the results may be available the same day that the sample is received in the testing laboratory. One day means results are available 1 day after the sample is received in the laboratory.

Same day/1 day

Maximum Laboratory Time Defines the maximum time from specimen receipt at Mayo Clinic Laboratories until the release of the test result

2 days

Performing Laboratory Location Indicates the location of the laboratory that performs the test

Rochester