TEST CATALOG ORDERING & RESULTS SPECIMEN HANDLING CUSTOMER SERVICE EDUCATION & INSIGHTS
Test Catalog

Test ID: COU    
Cobalt, 24 Hour, Urine

Useful For Suggests clinical disorders or settings where the test may be helpful

Detecting cobalt exposure

 

Monitoring metallic prosthetic implant wear

 

This test is not useful to assess vitamin B12 activity.

Clinical Information Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

Cobalt is rare but widely distributed in the environment. It is an essential cofactor in vitamin B12. While cobalt is an essential element, cobalt deficiency has not been reported in humans.

 

Cobalt is used in the manufacture of hard alloys with high melting points and resistance to oxidation. Cobalt salts are also used in the glass and pigment industry. Previously, cobalt salts were sometimes used as foam stabilizers in the brewing industry; this practice was banned due to the cardiovascular diseases it induced. The radioactive isotope of cobalt, (60)Co, is used as a gamma emitter in experimental biology, cancer therapy, and industrial radiography.

 

Cobalt is not highly toxic, but large doses will produce adverse clinical manifestations. Acute symptoms are pulmonary edema, allergy, nausea, vomiting, hemorrhage, and renal failure. Chronic symptoms include pulmonary syndrome, skin disorders, and thyroid abnormalities. The inhalation of dust during machining of cobalt alloyed metals can lead to interstitial lung disease. Improperly handled (60)Co can cause radiation poisoning from exposure to gamma radiation.

 

Urine cobalt concentrations are likely to be increased above the reference value in patients with metallic joint prosthesis. Prosthetic devices produced by Zimmer Company and Johnson and Johnson typically are made of aluminum, vanadium, and titanium. Prosthetic devices produced by Depuy Company, Dow Corning, Howmedica, LCS, PCA, Osteonics, Richards Company, Tricon, and Whiteside typically are made of chromium, cobalt, and molybdenum. This list of products is incomplete, and these products change occasionally; see prosthesis product information for each device for composition details.

Reference Values Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

0-17 years: not established

> or =18 years: 0.2-3.5 mcg/24 hours

Interpretation Provides information to assist in interpretation of the test results

Concentrations of 2.0 mcg/specimen or more indicate excess exposure. There are no Occupational Safety and Health Administration (OSHA) blood or urine criteria for occupational exposure to cobalt.

 

Prosthesis wear is known to result in increased circulating concentration of metal ions. In a patient with a cobalt-based implant, modest increase (2-4 mcg/specimen) in urine cobalt concentration is likely to be associated with a prosthetic device in good condition. Excessive exposure is indicated when urine cobalt concentration is above 5 mcg/specimen, consistent with prosthesis wear. Urine concentrations above 20 mcg/specimen in a patient with a cobalt-based implant suggest significant prosthesis wear. Increased urine trace element concentrations in the absence of corroborating clinical information do not independently predict prosthesis wear or failure.

Cautions Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

Specimen collection procedures for cobalt require special specimen collection tubes, rigorous attention to ultraclean specimen collection and handling procedures, and analysis in an ultraclean facility. Unless all of these precautions are taken, elevated urine cobalt results may be an incidental and misleading finding.

 

Metal-free urine collection procedures must be followed (see Trace Metals Analysis Specimen Collection and Transport in Special Instructions).

Clinical Reference Recommendations for in-depth reading of a clinical nature

1. Keegan GM, Learmonth ID, Case CP: A systematic comparison of the actual, potential, and theoretical health effects of cobalt and chromium from industry and surgical implants. Crit Rev Toxicol. 2008;38:645-674

2. Lhotka C, Szekes T, Stefan I, et al: Four-year study of cobalt and chromium blood levels in patients managed with two different metal-on-metal total hip replacements. J Orthop Res. 2003;21:189-195

3. Lison D, De Boeck M, Verougstraete V, Kirsch-Volders M: Update on the genotoxicity and carcinogenicity of cobalt compounds. Occup Environ Med. 2001;58(10):619-625

Special Instructions Library of PDFs including pertinent information and forms related to the test